Extraction of late perforated high voltage defibrillator lead utilizing intra cardiac echocardiography

Ahmed Al Fagih*, Yahia Al Hebaishi, Saleh Al Ghamdi, Khalid Dagriri, Adil Ahmed, Saeed Al Ahmari, Ali Mahboob, Ayman Al khadra
Department of Adult Cardiology, Prince Sultan Cardiac Centre, Riyadh, Kingdom of Saudi Arabia
DOI: 10.13055/ojcar_5_1_2.141008

Corresponding Address:
* Department of Adult Cardiology, Prince Sultan Cardiac Center, Riyadh, Kingdom of Saudi Arabia; Email: aafagih@yahoo.com; Phone: +96614777714 Ex. 8765; Fax: +96614778771


Late perforation of rhythm device lead is a well recognized complication of implanted devices with varying degree of clinical presentation from asymptomatic patients to sudden cardiac death. We report a patient with incidental finding of late right ventricular lead perforation during routine device follow up. Endovenous extraction of the lead was performed uneventfully under intracardiac echocardiography guidance with cardiac surgery team backup.

Keywords: Intra cardiac echo, Cardiac CT, Extraction, ICD.

© Al Fagih et al.; licensee Ross Science Publishers
open-access license: ROSS Open Access articles will be distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by-nc/3.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided that the original work will always be cited properly.

Article History:
Receiving date: 26-8-2014
Acceptance date: 8-10-2014
Electronic publication date: 8-10-2014
Year of Publication: 2014

Case description

53 year old lady with long standing history of ischemic heart disease and aborted sudden cardiac death underwent single chamber defibrillator implantation as a secondary prevention. The implanted lead was passive fixation, dual coil, 65 centimeters, 7Fr, (Riata®, model 1560, St. Jude Medical). Chest X-ray (CXR) post implantation showed apical right ventricular (RV) lead position. Loss of capture with high pacing impedance was noticed during first routine device follow up six weeks after implantation. CXR showed suspicious lead migration confirmed by fluoroscopy (Figure 1). Trans-thoracic echocardiography showed RV lead perforating the RV free wall and there was no evidence of pericardial effusion (Figure 2). Cardiac computed tomography (CT) with three dimensional reconstructions was performed to assess the exact site and the extent of perforation before attempting extraction (Figure 3). The procedure was performed under local anesthesia and conscious sedation with a cardiac surgical team standby. The perforating lead was explanted by simple traction under intra-cardiac echocardiography (ICE) guidance to check for immediate leak or temponade (Figure 4). After removal of the old lead, a new RV lead (Medtronic sprint quarto 6947, 65cm) was implanted in RV septum with good sensing and pacing parameters. There was no evidence of Immediate Post-operative pericardial effusion as well as before discharge as assessed by trans-thoracic echocardiography. Routine device check six weeks later showed normal pacing parameters.

Figure 1. Lateral(a), Left anterior obligue (b) and posteo-anterior (c) flouruscopy of the heart showed right ventricular lead migration and the tip is just anterior to the left hemidiaphragm.
Figure 2. Apical four chamber trans-thoracic echocardiography view showed right ventricular lead perforating the right ventricular free wall (Arrow) with no signs of temponade.
Figure 3. Cardiac CT with 3dimentional reconstruction demonstrating the site and the extent of the lead migration. Right ventricular lead is visible outside the outline of the heart (arrow).
Figure 4. Postero-anterior fluoroscopy oh the heart during extraction stages (a,b,c) under intra-cardiac echocardiography (ICE) monitoring. Notices the ICE probe I the right ventricle (arrow).


Delayed ventricular lead peroration is a rare but potentially life threatening complication of device implantation [1]. The clinical presentation of late perforation might vary widely from minimal symptoms to cardiac death [2, 3]. In our case, the patient was asymptomatic despite significant migration distance of the perforating lead. Although computed tomography might be necessary to delineate the site and the extent of perforation, the diagnosis can be confirmed by chest X-ray or echocardiography [1, 2]. Percutaneous lead removal can be performed with extreme caution under transthoracic and/or transesophageal echocardiography monitoring in operating room with cardiac surgery team backup [4-6]. Furthermore, defibrillator passive fixation lead extraction might carry more risk because of bulky lead tip which may cause tissue damage during removal. Open chest surgery might be considered as a safe alternative but it may be associated with longer hospital stay. Use of intra cardiac echocardiography during invasive cardiac procedures has been accepted widely with accurate imaging and less requirement of general anesthesia [7, 8]. We described in this case the feasibility of intra cardiac echocardiography as alternative and adjunctive tool to monitor for possible complications such as temponade during extraction procedure.

Conflict of Interests



[1] Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007; 30: 28–32. [PMID: 17241311] [DOI: 10.1111/j.1540-8159.2007.00575.x]
[2] Ramirez MF, Ching CK, Ho KL, Teo WS. “The attack of the 52 cm lead”: an unusual case of late cardiac perforation by a passive-fixation permanent pacemaker lead. Int J Cardiol. 2007; 115: e5–7. [PMID: 16901560] [DOI: 10.1016/j.ijcard.2006.05.064]
[3] Tziakas D, Alexoudis A, Konstantinou F, Chalikias G, Stakos D, Bougioukas G. A rare case of late right ventricular perforation by a passive-fixation permanent pacemaker lead. Europace. 2009; 11: 968–9. [PMID: 19443432] [DOI: 10.1093/europace/eup120]
[4] Laborderie J, Barandon L, Ploux S, Deplagne A, Mokrani B, Reuter S, Le Gal F, Jais P, Haissaguerre M, Clementy J, Bordachar P. Management of subacute and delayed right ventricular perforation with a pacing or mplantable cardioverter-defibrillator lead. Am J Cardiol. 2008; 102: 1352–5. [PMID: 18993154] [DOI: 10.1016/j.amjcard.2008.07.025]
[5] Khan MN, Joseph G, Khaykin Y, Ziada KM, Wilkoff BL. Delayed lead perforation: a disturbing trend. Pacing Clin Electrophysiol. 2005; 28: 251–3. [PMID: 15733190] [DOI: 10.1111/j.1540-8159.2005.40003.x]
[6] Satpathy R, Hee T, Esterbrooks D, Mohiuddin S. Delayed defibrillator lead perforation: an increasing phenomenon. Pacing Clin Electrophysiol. 2008; 31: 10–2. [PMID: 18181902]
[7] Macdonald ST, Newton JD, Ormerod OJ. Intracardiac echocardiography off piste? Closure of the left atrial appendage using ICE and local anesthesia.. Catheter Cardiovasc Interv. 2011; 77: 124–7. [PMID: 20517996] [DOI: 10.1002/ccd.22652]
[8] Kefer J, Astarci P, Renkin J. Intracardiac echo imaging of a transfemoral aortic Edwards SAPIEN valve implantation in a degenerated bioprosthesis (valve-in-valve). J Invasive Cardiol. 2010; 22: 341–2. [PMID: 20603509]
2014 Ross Science Publishers